beautypg.com

Burst frequency, Afg-2225 series user manual, Panel operation 1. press the burst key – GW Instek AFG-2225 User Manual User Manual

Page 120: Panel operation 1. press the freq/rate key, Press f1~f5 to select the frequency unit

background image

AFG-2225 Series User Manual

120

In Gated mode, burst count, burst cycle and trigger source are
ignored. If a trigger is input, then the trigger will be ignored and will
not generate any errors.

Panel Operation

1. Press the Burst key.



2. To select either N Cycle (F1)

or Gate (F2).

yl



~

at



Burst Frequency

In the N Cycle and Gated modes, the waveform frequency sets the
repetition rate of the burst waveforms. In N-Cycle mode, the burst is
output at the waveform frequency for the number of cycles set. In
Gated mode the waveform frequency is output while the trigger is
high. Burst mode supports sine, square or ramp waveforms.

Panel Operation

1. Press the FREQ/Rate key.

/

2. The FREQ parameter will become highlighted

in the parameter window.

3. Use the arrow keys and scroll

wheel or number pad to enter
the frequency.

/

4. Press F1~F5 to select the

frequency unit.

uz



~

z

