AFG-2225 Series User Manual

214

the output termination from high impedance to
50Ω will half the amplitude.
If the output termination is set to high impedance,
dBm units cannot be used. The units will default to
Vpp.

Syntax

OUTPut[1|2]:LOAD {DEFault|INFinity}

Example

OUTP1:LOAD DEF

Sets the channel 1 output termination to 50Ω.

Query Syntax

OUTPut[1|2]:LOAD?

Return Parameter DEF

Default

INF

INFinity

Example

OUTP1:LOAD?
DEF

The output termination for channel 1 is set to 50Ω.

SOURce[1|2]:VOLTage:UNIT

Source Specific
Command

Description

Sets or queries the output amplitude units. There
are three types of units: VPP, VRMS and DBM.

Note

The units set with the VOLTage:UNIT command
will be used as the default unit for all amplitude
units unless a different unit is specifically used for
a command.
If the output termination is set to high impedance,
dBm units cannot be used. The Units will
automatically default to Vpp.

Syntax

SOURce[1|2]:VOLTage:UNIT {VPP|VRMS|DBM}

Example

SOUR1:VOLT:UNIT VPP

Sets the amplitude units to Vpp.

Query Syntax

SOURce[1|2]:VOLTage:UNIT?

Return Parameter VPP

Vpp

VRMS

Vrms