beautypg.com

Source[1|2]:sweep:spacing, Source[1|2]:sweep:time – GW Instek AFG-2225 User Manual User Manual

Page 245

background image

REMOTE INTERFACE

245

SOURce[1|2]:SWEep:SPACing

Source Specific
Command

Description

Sets linear or logarithmic sweep spacing. The
default spacing is linear.

Syntax

SOURce[1|2]:SWEep:SPACing {LINear|LOGarithmic}

Example

SOUR1:SWE:SPAC LIN

Sets the spacing to linear.

Query Syntax

SOURce[1|2]:SWEep:SPACing?

Return Parameter LIN

Linear spacing

LOG

Logarithmic spacing

Example

SOUR1:SWE:SPAC?
LIN

The spacing is currently set as linear.

SOURce[1|2]:SWEep:TIME

Source Specific
Command

Description

Sets or queries the sweep time. The default sweep
time is 1 second.

Note

The function generator automatically determines
the number of frequency points that are used for
the sweep based on the sweep time.

Syntax

SOURce[1|2]:SWEep:TIME
{|MINimum|MAXimum}

Parameter

1 ms ~ 500 s

Example

SOUR1:SWE:TIME +1.0000E+00

Sets the sweep time to 1 second.

Query Syntax

SOURce[1|2]:SWEep:TIME? {|
MINimum|MAXimum}

Return Parameter

Returns sweep time in
seconds.