beautypg.com

Source[1|2]:marker:frequency, Source[1|2]:marker – GW Instek AFG-2225 User Manual User Manual

Page 247

background image

REMOTE INTERFACE

247

SOURce[1|2]:MARKer:FREQuency

Source Specific
Command

Description

Sets or queries the marker frequency. The default
marker frequency is 550 Hz. The marker frequency
is used to output a trigger out signal from the
trigger terminal on the rear panel.

Note

The marker frequency must be between the start
and stop frequencies. If the marker frequency is set
to a value that is out of the range, the marker
frequency will be set to the center frequency and a
“settings conflict” error will be generated.

Syntax

SOURce[1|2]:MARKer:FREQuency
{|MINimum|MAXimum}

Parameter

1μHz ~ 25 MHz
1 μHz ~ 1 MHz (Ramp)

Example

SOUR1:MARK:FREQ +1.0000E+03

Sets the marker frequency to 1 kHz.

Query Syntax

SOURce[1|2]:MARKer:FREQuency? [MINimum|
MAXimum]

Return Parameter

Returns the marker
frequency in Hz.

Example

SOUR1:MARK:FREQ? MAX
+1.0000E+03

Returns the marker frequency (1 kHz).

SOURce[1|2]:MARKer

Source Specific
Command

Description

Turns the marker frequency on or off. The default
is off.

Note

MARKer ON The SYNC signal goes logically

high at the start of each sweep and
goes low at the marker frequency.