beautypg.com

Source[1|2]:pm:internal:function, Source[1|2]:pm:internal:frequency – GW Instek AFG-2225 User Manual User Manual

Page 234

background image

AFG-2225 Series User Manual

234

Example

SOUR1:PM:SOUR?
INT

The PM source is set to internal.

SOURce[1|2]:PM:INTernal:FUNction

Source Specific
Command

Description

Sets the shape of the modulating waveform from
sine, square, triangle, upramp and dnramp. The
default shape is sine.

Note

Square and triangle waveforms have a 50% duty
cycle. Upramp and dnramp have a symmetry to
100% and 0%, respectively. .

Syntax

SOURce[1|2]:PM:INTernal:FUNction
{SINusoid|SQUare|TRIangle|UPRamp|DNRamp}

Example

SOUR1:PM:INT:FUN SIN

Sets the PM modulating wave shape to sine. .

Query Syntax

SOURce[1|2]:PM:INTernal:FUNction?

Return Parameter SIN

Sine

UPRAMP

Upramp

SQU

Square

DNRAMP

Dnramp

TRI

Triangle

Example

SOUR1:PM:INT:FUNC?
SIN

The shape for the modulating waveform is Sine.

SOURce[1|2]:PM:INTernal:FREQuency

Source Specific
Command

Description

Sets the modulating waveform frequency for
internal sources. The default frequency is set to
100Hz.

Syntax

SOURce[1|2]:PM:INTernal:FREQuency
{|MINimum|MAXimum}

Parameter

2 mHz~ 20 kHz