Output commands, Source[1|2]:function – GW Instek AFG-2225 User Manual User Manual
Page 206

AFG-2225 Series User Manual
206
Output Commands
Unlike the Apply commands, the Output commands are low level 
commands to program the function generator. 
This section describes the low-level commands used to program the 
function generator. Although the APPLy command provides the 
most straightforward method to program the function generator, the 
low-level commands give you more flexibility to change individual 
parameters. 
SOURce[1|2]:FUNCtion
Source Specific 
Command 
Description
The FUNCtion command selects and outputs the 
selected output. The User parameter outputs an 
arbitrary waveform previously set by the 
SOURce[1|2]:FUNC:USER command. 
Note
If the function mode is changed and the current 
frequency setting is not supported by the new 
mode, the frequency setting will be altered to next 
highest value. 
Vpp and Vrms or dBm amplitude values may have 
different maximum values due to differences such 
as crest factor. For example, if a 5Vrms square 
wave is changed to a sinewave, then the Vrms is 
automatically adjusted to 3.536. 
The modulation, burst and sweep modes can only 
be used with some of the basic waveforms. If a 
mode is not supported, the conflicting mode will 
be disabled. See the table below. 