beautypg.com

GW Instek AFG-2225 User Manual User Manual

Page 284

background image

AFG-2225 Series User Manual

284

-313 Calibration memory lost;memory corruption detected

Indicates that a fault (check sum error) has occurred
with the non-volatile memory that stores the
calibration data.

-314 Save/recall memory lost;memory corruption detected

Indicates that a fault (check sum error) has occurred
with the non-volatile memory that stores the
save/recall files.

-315 Configuration memory lost;memory corruption detected

Indicates that a fault (check sum error) has occurred
with the non-volatile memory that stores the
configuration settings.

-350 Queue overflow

Indicates that the error queue is full (over 20
messages generated, and not yet read). No more
messages will be stored until the queue is empty.
The queue can be cleared by reading each message,
using the *CLS command or restarting the function
generator.