beautypg.com

Burst cycle/burst count – GW Instek AFG-2225 User Manual User Manual

Page 121

background image

MODULATION

121

Range

Frequency

1uHz~25MHz

Freqency – Ramp

1uHz~1MHz

Default

1kHz

Note

Waveform frequency and burst period are not the
same. The burst period is the time between the
bursts in N-Cycle mode.

Burst Cycle/Burst Count

The burst cycle (burst count) is used to define the number of cycles
that are output for a burst waveform. Burst cycle is only used with
N-cycle mode (internal, external or manual source). The default
burst cycle is 1.

Panel Operation

1. Press the Burst key.



2. Press F1 (N Cycle).

yl



3. Press F1 (Cycles).

yls



4. The Cycles parameter will become highlighted

in the Waveform Display area.