beautypg.com

Source[1|2]:sweep:state – GW Instek AFG-2225 User Manual User Manual

Page 241

background image

REMOTE INTERFACE

241

Select Sweep

Time

Select the sweep

trigger source

Select the marker

frequency

5. Choose the sweep time using the

SOURce[1|2]:SWE:TIME command.

6. Select an internal or external sweep trigger

source using the SOURce[1|2]:SOUR
command.

7. To output a marker frequency from the

trigger out, use The
SOURce[1|2]:MARK:FREQ command. To
enable marker frequency output, use the
SOURce[1|2]:MARK ON command.

The marker frequency can be set to a value within
the sweep span.

SOURce[1|2]:SWEep:STATe

Source Specific
Command

Description

Sets or disables Sweep mode. By default Sweep is
disabled. Sweep modulation must be enabled
before setting other parameters.

Note

Any modulation modes or Burst mode will be
disabled if sweep mode is enabled.

Syntax

SOURce[1|2]:SWEep:STATe {OFF|ON}

Example

SOUR1:SWE:STAT ON

Enables sweep mode.

Query Syntax

SOURce[1|2]:SWEep:STATe?

Return Parameter 0

Disabled (OFF)

1

Enabled (ON)

Example

SOUR1:SWE:STAT?
1

Sweep mode is currently enabled.