REMOTE INTERFACE

271

Save and Recall Commands

Up to 10 different instrument states can be stored to non-volatile
memory (memory locations 0~9).

*SAV

Instrument
Command

Description

Saves the current instrument state to a specified
save slot. When a state is saved, all the current
instrument settings, functions and waveforms are
also saved.

Note

The *SAV command doesn’t save waveforms in
non-volatile memory, only the instrument state.
The *RST command will not delete saved
instrument states from memory.

Syntax

*SAV {0|1|2|3|4|5|6|7|8|9}

Example

*SAV 0

Save the instrument state to memory location 0.

*RCL

Instrument
Command

Description

Recall previously saved instrument states from
memory locations 0~9.

Syntax

*RCL {0|1|2|3|4|5|6|7|8|9}

Example

*RCL 0

Recall instrument state from memory location 0.

MEMory:STATe:DELete

Instrument
Command

Description

Delete memory from a specified memory location.

Syntax

MEMory:STATe:DELete {0|1|2|3|4|5|6|7|8|9}

Example

MEM:STAT:DEL 0