Syntax

SOURce[1|2]:APPLy:USER [ [,
[,] ]]

Parameter

1μHz~60MHz

1mVpp~10Vpp (50Ω)

-4.99V~4.99V (50Ω)

Example

SOUR1:APPL:USER

SOURce[1|2]:APPLy?

Source Specific
Command

Description

Outputs a string with the current settings.

Note

The string can be passed back appended to the
Apply Command.

Syntax

SOURce[1|2]:APPLy?

Return Parameter

Function, frequency,
amplitude, offset

Example

SOUR1:APPL?
SIN +5.0000000000000E+03,+3.0000E+00,-2.50E+00

Returns a string with the current function and