beautypg.com

Source[1|2]:dcoffset – GW Instek AFG-2225 User Manual User Manual

Page 210

background image

AFG-2225 Series User Manual

210

Syntax

SOURce[1|2]:AMPlitude {< amplitude>
|MINimum|MAXimum}

Example

SOUR1:AMP MAX

Sets the amplitude to the maximum for the current
mode.

Query Syntax

SOURce[1|2]:AMPlitude? {MINimum|MAXimum}

Return Parameter

Returns the amplitude for
the current mode.

Example

SOUR1:AMP? MAX

+5.0000E+00

The maximum amplitude that can be set for the
current function is 5 volts.

SOURce[1|2]:DCOffset

Source Specific
Command

Description

Sets or queries the DC offset for the current mode.

Note

The offset parameter can be set to MINimum,
MAXimum or DEFault. The default offset is 0
volts. The offset is limited by the output amplitude
as shown below.

|Voffset| < Vmax – Vpp/2

If the output specified is out of range, the
maximum offset will be set.
The offset is also determined by the output
termination (50Ω or high impedance). If the offset
has been set and the output termination has
changed from 50Ω to high impedance, the offset
will double. Changing the output termination from
high impedance to 50Ω will half the offset.