Pulse configuration commands, Source[1|2]:pulse:period – GW Instek AFG-2225 User Manual User Manual
Page 216
AFG-2225 Series User Manual
216
Pulse Configuration Commands
The pulse chapter is used to control and output pulse waveforms.
Unlike the APPLy command, low level control is possible including
setting the rise time, fall time, period and pulse width.
Fall time
Rise time
Pulse Width
Period
50%
90%
10%
90%
50%
10%
SOURce[1|2]:PULSe:PERiod
Source Specific
Command
Description
Sets or queries the pulse period. The default period
is 1 ms.
Note
The pulse period must be greater than the pulse
width and edge time(1.6x) combined.
Pulse Width + (1.6 * Edge Time) < Period
If the edge time or pulse width are too great, they
will automatically be reduced to fit the period by
the function generator.
The PULSe:PERiod function will change the period
for all functions, not just for the pulse waveforms.
If a different function is chosen and the current
period is out of range, the period will be
automatically adjusted to suit the new function.
Syntax
SOURce[1|2]:PULSe:PERiod
{
Example
SOUR1:PULS:PER MIN
Sets the period to the minimum time allowed.
Query Syntax
SOURce[1|2]:PULSe:PERiod? [MINimum|MAXimum]
Return Parameter
40ns~2000s