Example

SOUR1:BURS:MODE?
TRIG

The current burst mode is triggered.

SOURce[1|2]:BURSt:NCYCles

Source Specific
Command

Description

Sets or queries the number of cycles (burst count)
in triggered burst mode. The default number of
cycles is 1. The burst count is ignored in gated
mode.

Note

If the trigger source is set to immediate, the
product of the burst period and waveform
frequency must be greater than the burst count:
Burst Period X Waveform frequency > burst count
If the burst count is too large, the burst period will
automatically be increased and a “Settings
conflict” error will be generated.
Only sine and square waves are allowed infinite
burst above 25 MHz.

Syntax

SOURce[1|2]:BURSt:NCYCles{< # cycles>
|INFinity|MINimum |MAXimum}

Parameter

<# cycles>

1~65535 cycles.

INFinity

Sets the number to continuous.

MINimum

Sets the number to minimum allowed.

MAXimum

Sets the number to maximum allowed.

Example

SOUR1:BURS:NCYCl INF

Sets the number of burst cycles to continuous
(infinite).

Query Syntax

SOURce[1|2]:BURSt:NCYCles? [MINimum|MAXimum]

Return Parameter

Returns the number of cycles.

INF

INF is returned if the number of cycles
is continuous.