AFG-2225 Series User Manual

182

Remote control terminal connection

Terminal
application

Invoke the terminal application such as MTTTY
(Multi-Threaded TTY). For USB, set the COM port,
baud rate, stop bit, data bit, and parity
accordingly.
To check the COM port No, see the Device
Manager in the PC. For WinXP, Control panel →
System → Hardware tab.

Functionality
check

Run this query command via the terminal.
*idn?
This should return the Manufacturer, Model
number, Serial number, and Firmware version in
the following format.
GW INSTEK, AFG-2225, SN:XXXXXXXX,Vm.mm

Note: ^j or ^m can be used as the terminal
character when using a terminal program.

PC Software

The proprietary PC software, downloadable from
GWInstek website, can be used for remote control.

Display

When a remote connection is established all panel
keys are locked bar F5.

1. Press REM/LOCK (F5) to

return the function generator
to local mode.

K

