beautypg.com

GW Instek AFG-2225 User Manual User Manual

Page 208

background image

AFG-2225 Series User Manual

208

Noise

Not applicable

User

1μHz~60MHz

If the function mode is changed and the current
frequency setting is not supported by the new
mode, the frequency setting will be altered to next
highest value.
The duty cycle of square waveforms depends on the
frequency settings.
1.0% to 99.0%(frequency≤100 KHz)
10% to 90% (100 KHz ≤ frequency ≤1MHz)
50% (frequency ≤ 25 MHz)
If the frequency is changed and the set duty cycle
cannot support the new frequency, the highest
duty cycle available at that frequency will be used.
A “settings conflict” error will result from the
above scenario.

Syntax

SOURce[1|2]:FREQuency
{|MINimum|MAXimum}

Example

SOUR1:FREQ MAX

Sets the frequency to the maximum for the current
mode.

Query Syntax

SOURce[1|2]:FREQuency?

Return Parameter

Returns the frequency for
the current mode.

Example

SOUR1:FREQ? MAX

+1.0000000000000E+06

The maximum frequency that can be set for the
current function is 1MHz.