beautypg.com

Source[1|2]:apply:sinusoid, Source[1|2]:apply:square – GW Instek AFG-2225 User Manual User Manual

Page 202

background image

AFG-2225 Series User Manual

202

The offset is also determined by the output
termination (50Ω or high impedance). If the offset
has been set and the output termination has
changed from 50Ω to high impedance, the offset
will double. Changing the output termination from
high impedance to 50Ω will half the offset.

SOURce[1|2]:APPLy:SINusoid

Source Specific
Command

Description

Outputs a sine wave from the selected channel
when the command has executed. Frequency,
amplitude and offset can also be set.

Syntax

SOURce[1|2]:APPLy:SINusoid [
[, [,] ]]

Parameter

1μHz~25MHz

1mVpp~10Vpp (50Ω)
(3.536 Vrms)

-4.99V~4.99V (50Ω)

Example

SOUR1:APPL:SIN 2KHZ,MAX,MAX

Sets frequency to 2kHz and sets the amplitude and
offset to the maximum.

SOURce[1|2]:APPLy:SQUare

Source Specific
Command

Description

Outputs a square wave from the selected channel
when the command has executed. Frequency,
amplitude and offset can also be set. The duty
cycle is set to 50%.

Syntax

SOURce[1|2]:APPLy:SQUare [
[, [,] ]]

Parameter

1μHz~25MHz

1mVpp~10Vpp (50Ω)

-4.99V~4.99V (50Ω)

Example

SOUR1:APPL:SQU 2KHZ,MAX,MAX