beautypg.com

Error messages in the message log – Luminex 100 IS User Manual, Version 2.1 User Manual

Page 93

background image

x

MAP Technology

Software Tour

PN 89-00002-00-061 Rev. A

7 - 33

and commands that the system has performed as well as each
operation’s progress, time and date, and results.

Figure 53. Message Log

You also see a message in the log if multiple microtiter plates have
been used during a batch.

The log also displays actions in color-coded text and shading. Items
in the message log appear in the following color code:

Green text represents a successful system calibration,
verification command, an acquisition, or maintenance functions.

Red text represents failed commands or errors.

Black text represents normal processes and actions.

Yellow shading behind the text represents that a detailed
description about the processes or actions is available. This color
may vary from system to system depending on your tool tip color
system settings.
This color shading appears only when a more detailed
explanation exists for the item.

To see this detailed description, double-click the row that is
shaded in yellow. A dialog box appears providing details about
that action or process. This function is particularly helpful if you
need to diagnose a problem with the system or reagents.

If an error occurs during the batch, it is recorded in the message log.
See the subsection below this paragraph for more information about
errors.

Error Messages in the
Message Log

If an error occurs within your operations, click the diagnostics tab
and look at the message log. The error will appear as a row with red
text. If the row with the red text has a yellow background, double