Luminex 100 IS User Manual, Version 2.1 User Manual
Page 67
x
MAP Technology
Software Tour
PN 89-00002-00-061 Rev. A
7 - 7
The following table describes the color coding with relation to types
of message received in the status bar.
Table 1. Status Bar Color Coding
Category
Indicates
Command
State
Indicates communication with the analyzer
or operations being processed
Green
Idle or processing
Yellow
Connecting, pausing, or paused
Red
Disconnected or locked out
Device Status
Indicates the current process of or warning
about the analyzer
Green
Running or standby
Yellow
Busy, pressurizing, sheath is empty, or warming
up
Red
Not ready or disconnected
Device
Activity
Indicates the activity the Analyzer is
performing
Note:
The device activity has no color code.
Possible
activities
include:
•
Idling: waiting for a command
•
Aspirating: drawing in sample
•
Collecting Data: collecting data
•
Sanitizing: sanitizing the instrument
•
Washing: washing the instrument
•
Priming: priming the instrument
•
Calibrating: calibrating the instrument
•
Canceling: canceling a command
•
Backflushing: backflushing the instrument
•
Draining: draining the instrument
•
Warming Up: warming up the instrument
•
Verifying: verifying calibration
•
Self Diagnosing: performing self diagnostic
routine
•
Soaking: soaking the probe
Laser Status
Laser temperature and readiness status
Green
Lasers warmed up
Yellow
Warmup timer countdown
Red
Lasers off