beautypg.com

Luminex 100 IS User Manual, Version 2.1 User Manual

Page 55

background image

x

MAP Technology

System Calibration

PN 89-00002-00-061 Rev. A

6 - 3

Figure 21. Update Calibration Targets Dialog Box

10. Enter the values listed from the Certificates of Analysis included

with your calibrators into the CAL1 and CAL2 entry fields.

11. Click OK. The dialog box disappears.

12. Ensure that the IS is set to draw the CAL1 and CAL2 beads from

the wells where you loaded them (in step 2).

13. Click CAL1 from the Maintenance tab. A Confirmation Screen

dialog box appears.

14. Click OK and wait until CAL1 completes. The device status

section in the status bar changes from “Running” to “Standby”.

15. Click CAL2 from the Maintenance tab. A Confirmation Screen

dialog box appears.

16. Click OK and wait until CAL2 completes. The device status

section in the status bar changes from “Running” to “Standby”.

You must run system controls following calibration. Continue on to
the “Run System Controls” on page 6-4.

Note:

Ensure that you enter

the correct target values for
each parameter before clicking
OK. If necessary, re-enter the
information to correct any
incorrect information.

Note:

The message log on the

Diagnostics tab shows the
status of the CAL1 and CAL2
commands. If an error occurs
during system calibration, the
Diagnostics tab turns red.