System monitor – Luminex 100 IS User Manual, Version 2.1 User Manual

Page 91

background image

x

MAP Technology

Software Tour

PN 89-00002-00-061 Rev. A

7 - 31

These tools help you find information about the system and what
happens during sample acquisition, data reduction, and other
functions.

For instance, if you return to the system to run an operation after
taking a break, you can look at the Diagnostics tab at the system
monitor to verify if you need to recalibrate.

System Monitor

The system monitor provides information on the status of the system
and various components within the system.

Figure 51. System Monitor

Table 4 lists representative values. Your values may vary from those
shown.

Table 4. System Monitor Values

Property

Value

Air Pressure

7.5 ± 1.5 PSI

Sheath Pressure

7 PSI

Delta Cal Temperature

0°C

Board Temperature

30°C

CL1 Temperature

25°C

CL2 Temperature

25°C

XYP Instrument Board
Temperature

40°C

XYP Instrument Heater
Temperature

45°C

XYP Instrument Heater
Temperature In Range

Yes