System monitor – Luminex 100 IS User Manual, Version 2.1 User Manual
Page 91
x
MAP Technology
Software Tour
PN 89-00002-00-061 Rev. A
7 - 31
These tools help you find information about the system and what
happens during sample acquisition, data reduction, and other
functions.
For instance, if you return to the system to run an operation after
taking a break, you can look at the Diagnostics tab at the system
monitor to verify if you need to recalibrate.
System Monitor
The system monitor provides information on the status of the system
and various components within the system.
Figure 51. System Monitor
Table 4 lists representative values. Your values may vary from those
shown.
Table 4. System Monitor Values
Property
Value
Air Pressure
7.5 ± 1.5 PSI
Sheath Pressure
7 PSI
Delta Cal Temperature
0°C
Board Temperature
30°C
CL1 Temperature
25°C
CL2 Temperature
25°C
XYP Instrument Board
Temperature
40°C
XYP Instrument Heater
Temperature
45°C
XYP Instrument Heater
Temperature In Range
Yes