100 IS User Manual Version 2.1

x

MAP Technology

8 - 4

PN 89-00002-00-061 Rev. A

Figure 88. Grasping the Tubing

4. Remove the filter with one hand, and hold the tubing with the

other hand.

5. Connect a new filter to the tubing and position the filter inside

the panel.

6. Reattach the panel door to the unit.

7. Plug in and power on the Luminex 100 analyzer.

XYP Instrument Air Intake
Filter

‹

To replace the air intake filter:

1. Disconnect the XYP instrument from ac power by turning off

the power switch on the rear of the XYP instrument, then
unplugging the XYP instrument power cord from the wall
source.

2. On the back of the XYP instrument, left side, gently remove the

screen from the XYP instrument filter.

Note:

Do not remove the

screws.