Install or replace the xyp instrument sample probe – Luminex 100 IS User Manual, Version 2.1 User Manual
Page 31
x
MAP Technology
Installation
PN 89-00002-00-061 Rev. A
4 - 5
Install or Replace the XYP
Instrument Sample Probe
To install the XYP instrument Sample Probe, which is longer
than the probe shipped with the system, follow these steps:
1. Make sure the Luminex 100 analyzer power switch is off and
that the XYP instrument and the Luminex 100 analyzer power
cords are unplugged from the outlet.
2. Remove the light housing directly above the Sample Arm by
grasping and firmly pulling out. The housing remains attached
by a wire. Place the housing on top of the analyzer.
Figure 6. Light Housing
Light housing