4 - 5

Install or Replace the XYP
Instrument Sample Probe

‹

To install the XYP instrument Sample Probe, which is longer

than the probe shipped with the system, follow these steps:

1. Make sure the Luminex 100 analyzer power switch is off and

that the XYP instrument and the Luminex 100 analyzer power
cords are unplugged from the outlet.

2. Remove the light housing directly above the Sample Arm by

grasping and firmly pulling out. The housing remains attached
by a wire. Place the housing on top of the analyzer.

Figure 6. Light Housing

Light housing