Installation

Assemble and

Power the Luminex

100 IS System

Before installing the system, ensure that the facility complies with all
system and safety requirements. Read the safety information that
begins in

“Safety Precautions” on page 2-2

. Try to position the

instrument to minimize temperature fluctuations.

Hardware Setup

The following sections describe how to set up the hardware
components of the Luminex 100 IS.

Connect the

Luminex 100

Analyzer and XYP

Instrument

‹

To assemble and power on the system:

1. Unpack the Luminex 100 analyzer, the XYP instrument, and the

PC and monitor. Review “Hardware” on page 5-2 and identify
each system component.

2. Place the XYP instrument on a clean, flat surface.

Figure 1. Red Shipping Pin

3. Remove the red shipping pin. Leave the silver knob in the XYP

instrument.

4. Attach the power cord to the power input module of the XYP

instrument and attach the communication cable to the XYP

4

Red shipping pin

Note:

Due to the weight of the

Luminex 100 analyzer, two people
are required for lifting.