Initialization – Luminex 100 IS User Manual, Version 2.1 User Manual

Page 37

background image

x

MAP Technology

Installation

PN 89-00002-00-061 Rev. A

4 - 11

5. Use the 3/32 inch hexagonal allen wrench to loosen the height

adjustment locking screw. See Figure 12.

6. Click the Maintenance tab.

7. Click the Sample Probe Down button to lower the Sample

Probe.

8. Using the thumb wheel, lower the probe until it is just touching

the top of the alignment discs or sphere.

9. Use a 3/32 inch hexagonal allen wrench to tighten the height

adjustment locking screw.

10. Click Sample Probe Up to raise the sampling probe.

11. Replace the clear plastic shield that covers the sampling probe

area.

Initialization

Initialize the PC

Upon powering your system, the PC runs through several self-
diagnostic tests to verify that the computer system is set up to run the
system software application.

‹

To initialize the PC:

1. Turn on the power button on the back of the Luminex 100

analyzer and turn on the PC. If you have these devices plugged
into a UPS or surge suppressor protector rather than directly into
the wall outlet, turn on the surge protection power strip power
switch.

2. The PC and system begin initializing automatically.

If the system initializes successfully, the software application
launches. If the system does not run through its initial tests
successfully, the software application does not launch and a warning
appears on your PC’s monitor screen.

The PC runs through the following tests during initialization:

RAM

ROM

Communications with hard drive

Communications with mouse