100 IS S Users Manual Version 2.1

x

MAP Technology

7 - 26

PN 89-00002-00-061 Rev. A

Run Self

Diagnostics

Run self diagnostics to see if the device and all system operations are
running correctly. The system tests these functions when running self
diagnostics:

Flash memory test

Microcontroller RAM test

Nonvolatile memory test

DSP program CRC test

DSP capture test

High voltage module

Channel backgrounds test

Pressurization test

Actuator test

Syringe pump test

Backflush valve test

‹

To run self diagnostics:

1. Click the Self-Diagnostics button from the maintenance tab. A

confirmation dialog box appears.

2. Click OK. The system begins processing the various self

diagnostic tests. As it completes the tests, the status bar changes
from a “Processing” command state to an “Idle” command state.

The self-diagnostic test should take less than one minute to
complete. If the self diagnosis failed, the system lets you obtain
more detailed information regarding the results of the self-
diagnostic test.