Luminex 100 IS User Manual, Version 2.1 User Manual
Page 86
100 IS S Users Manual Version 2.1
x
MAP Technology
7 - 26
PN 89-00002-00-061 Rev. A
Run Self
Diagnostics
Run self diagnostics to see if the device and all system operations are
running correctly. The system tests these functions when running self
diagnostics:
•
Flash memory test
•
Microcontroller RAM test
•
Nonvolatile memory test
•
DSP program CRC test
•
DSP capture test
•
High voltage module
•
Channel backgrounds test
•
Pressurization test
•
Actuator test
•
Syringe pump test
•
Backflush valve test
To run self diagnostics:
1. Click the Self-Diagnostics button from the maintenance tab. A
confirmation dialog box appears.
2. Click OK. The system begins processing the various self
diagnostic tests. As it completes the tests, the status bar changes
from a “Processing” command state to an “Idle” command state.
The self-diagnostic test should take less than one minute to
complete. If the self diagnosis failed, the system lets you obtain
more detailed information regarding the results of the self-
diagnostic test.