Luminex 100 analyzer ventilation filter – Luminex 100 IS User Manual, Version 2.1 User Manual
Page 139
x
MAP Technology
Maintenance and Cleaning
PN 89-00002-00-061 Rev. A
8 - 7
10. Hand-tighten the screw on the syringe arm.
11. Plug in the power cord and turn the analyzer on.
12. Prime the system twice, watching for any leaks in the syringe
area.
13. When the prime finishes, close the center door.
Luminex 100 Analyzer
Ventilation Filter
Clean the filter only when soiled.
To check the Luminex 100 analyzer ventilation filter, follow
these steps:
1. Disconnect the instrument from ac power by turning off the
power switch on the rear of the instrument, then unplugging the
instrument power cord from the wall source.
2. On the lower right side of the Luminex 100 analyzer, firmly push
the filter to the left.
3. Remove the filter from the left side of the Luminex 100 analyzer.
Figure 92. Luminex 100 Analyzer Ventilation Filter
4. Clean the filter with a vacuum or with distilled water. Stand the
filter upright to air dry.
5. Re-install it with the arrows facing up. The filter should click
into place.
6. Plug in and power on the Luminex 100 analyzer.
Luminex 100 analyzer ventilation filter (on bottom)