Required laboratory reagents, Software, Luminex 100 is performance specifications – Luminex 100 IS User Manual, Version 2.1 User Manual

Page 41

background image

x

MAP Technology

The System

PN 89-00002-00-061 Rev. A

5 - 3

Required

Laboratory

Reagents

Household bleach

70% isopropanol

Mild detergent

Software

Luminex 100 IS software provides complete control of the system
and performs data analysis. Your Luminex 100 IS is preloaded with
the Luminex 100 IS software. However, we supply a software CD in
case you need to reinstall the software.

This software requires a dedicated system. Unauthorized additional
software is prohibited and may result in improper operation of the
system.

Luminex 100 IS

Performance

Specifications

Speed

Speed: greater than or equal to 933 MHz Pentium III processor
with 256 MB RAM

USB communications link for fast data transfer

Automatic transfer of assay templates and new reagent
information into the system via a 3½” floppy disk or large
capacity read/write CD

Installation < 4 hours

System calibration < 10 minutes

System controls < 10 minutes

Barcode reader entry of sample IDs

Automatic post-analysis

Analyze one 96-well plates/hour depending on manufacturer’s
kit

Up to 100 xMAP microsphere sets per sample

Sheath flowrate 90 µL/sec ± 5 µL

Sample injection rate into detector area 1 µL/sec ± 0.05 µL

System warmup 30 minutes: Systems that remain inactive for at
least four hours will require a warm-up to restart the lasers. After
acquiring sample, running system calibrators, running system
controls, and warming up the instrument, the system resets the
four-hour internal clock.