100 IS S Users Manual Version 2.1

x

MAP Technology

7 - 28

PN 89-00002-00-061 Rev. A

1. Click New CAL Targ. from the Maintenance tab. An Update

Cal Targets dialog box appears.

Figure 48. Update Calibration Targets Dialog Box

Enter the values listed in the kit manufacturer’s instructions or
Certificate of Analysis included with your kit into the CAL1 and
CAL2 entry fields.

2. Click OK. The database saves the updated calibration and

control values.

Note:

Ensure that the correct

target values are added for
each parameter before clicking
OK. If necessary, re-enter the
information to correct any
incorrect information.