beautypg.com

Luminex 100 IS User Manual, Version 2.1 User Manual

Page 62

background image

100 IS S Users Manual Version 2.1

x

MAP Technology

7 - 2

PN 89-00002-00-061 Rev. A

The calibration and control target values can be found on the
certificate of analysis included in the shipment of products.

After the initial splash screen disappears, the system prompts you to
enter calibration and control target values.

The system stores the information you enter into the database and
proceeds to the main software screen.

When entering information into the system, you may see an asterisk
(*) at the top-right corner of the entry field. The asterisk indicates
that this information is required. If a field does not have an asterisk
by it, then that information is optional and not required.

Figure 29. Entry Field With Asterisk Example

Enter xMAP

Calibration and

Control Targets

For the First Time

A dialog box appears telling you that you must enter xMAP
calibration and control target values if you have never entered these
values into the system.

‹

To enter xMAP calibration and control targets:

1. Click OK on the initial dialog box to enter these target values.

2. Enter the lot number, expiration date, and target values for CAL1

and CAL2. This area includes an optional area in which you may
enter comments regarding the calibration and control lots.

3. Enter the lot number, expiration date, and target values for

CON1 and CON2.

4. Click OK. The Update CON Targets dialog box closes. The

main screen of the IS software appears.