100 IS S Users Manual Version 2.1

x

MAP Technology

7 - 40

PN 89-00002-00-061 Rev. A

Figure 62. Luminex Batch Setup Dialog Box

4. Enter the batch name (if different from the system default name

provided), a description (optional), and the creator name.

5. If you want to add samples for processing in addition to the

template commands in the batch you are creating, scroll down to
the last item on the command list section (to the left of the
Finish, Cancel, Load Patient List, and Help buttons). If you are
running a maintenance template, go to step 11 to save and finish
creating the new batch.

6. Click the Sample ID field on the last row for the empty well on

the microtiter plate.