Luminex 100 IS User Manual, Version 2.1 User Manual

Page 85

background image

x

MAP Technology

Software Tour

PN 89-00002-00-061 Rev. A

7 - 25

‹

To soak the probe:

1. Click Soak. A confirmation dialog box appears.

Figure 46. Confirmation Screen Dialog Box

2. Click the drop-down arrow to the right of the Eject/Retract

button. An image of the microtiter plate and reservoir appears in
the plate holder.

Figure 47. Microtiter Plate Image For Location Selection

3. Select a location from which to draw fluid using the image of the

microtiter plate and reservoir, and click Eject/Retract. The plate
holder ejects.

Click the X button at the top-right corner of the screen to close
the image of the microtiter plate. The image disappears.

4. Verify that the location you indicated is empty, place the

distilled water in the location, and click Eject/Retract to retract
the plate holder.

5. Click OK to continue. The command list on the Run Batch tab

indicates that it is soaking. Wait until the soak command
completes before initiating another command. Upon completion,
the command list indicates whether the command succeeded
with green text or the command failed with red text.

Note:

The system may show a

default location from which to
draw fluid. In this case, the
microtiter plate shows that well
with a yellow line inside the
well.