x

MAP Technology

Maintenance and Cleaning

PN 89-00002-00-061 Rev. A

8 - 3

2. Wipe all exterior surfaces with mild detergent, followed by a

10% bleach solution.

3. Open both doors of the system and clean all accessible surfaces

with detergent followed by a 10% bleach solution.

4. Dry the sheet metal surfaces to prevent corrosion.

5. Plug in and power on the Luminex 100 analyzer and XYP

instruments.

6. Calibrate the system.

Calibration and System
Controls

Run calibration and system controls at least once a month during
routine use and when:

following installation

if the system is powered off

if the system is moved

if a part is replaced

if the delta calibration temperature shown on the system monitor
(located on the Diagnostics tab) is more than ±3 degrees.

Each step usually takes less than one minute. You must run
LabMAP controls after each calibration
. See “Run System
Controls” on page 6-4.

Semi-Annually

Luminex 100 Analyzer Air
Intake Filter

‹

To replace the air intake filter:

1. Disconnect the instrument from ac power by turning off the

power switch on the rear of the instrument, then unplugging
the instrument power cord from the wall source.

2. On the back of the Luminex 100 analyzer, in the upper left

corner, remove the screw at the top of the panel and open the
panel door.

3. Grasp the tubing and pull the filter 3-4 inches from the unit.

Note:

Hold on to the tubing! Do

not allow the tubing to fall inside the
instrument.