Xyp instrument, Pc specifications, Electronics – Luminex 100 IS User Manual, Version 2.1 User Manual
Page 44: General
Luminex 100 IS User Manual Version 2.1
x
MAP Technology
5 - 6
PN 89-00002-00-061 Rev. A
Electronics
•
Reporter channel detection: A/D resolution 14 bits
•
Communications interface: USB
•
XYP instrument, communications interface: RS 232
XYP Instrument
General
•
Ambient temperature: 15°C to 30°C (59°F to 86°F)
•
Humidity: 20% to 80%, noncondensing
•
Altitude: operation up to 2400 m (7874 ft) above mean sea level
•
Physical dimensions: 44 cm (17.25 inches) W x 60 cm (23.5
inches) D x 8 cm (3 inches) H
•
Weight: 15 kg (33 lbs.)
•
UL installation category: UL Installation Category II, as defined
in Annex J of UL 61010A-1
•
Pollution degree: UL Pollution Degree 2, as defined in Section
3.7.3.2 of UL 61010A-1
•
Heater operating range: 35°C to 55°C (95°F to 131°F) with
tolerance 0°C to +2°C
•
XY Platform, input voltage range: 100 - 240 V~, 1.8 Amps,
47 - 63 Hz
•
AC inlet fuse: 3 Amp, 250 Vac, fast acting
PC Specifications
•
Computer: a Pentium
®
III or equivalent class PC, minimum
speed 933 MHz
•
Main memory: minimum of 256 MB
•
Hard disk drive: 10 GB minimum storage capacity
•
Communications, serial ports: Minimum of 2 RS-232 compatible
serial ports
•
Communications, parallel pointing device and serial ports:
Minimum of 1 parallel, 1 PS/2 compatible pointing device, and 2
RS-232 compatible serial ports
•
Communications, USB port: Minimum of one USB Version 1.1
compatible high speed port
•
Read, write removable media: CDR/W-type drive, standard 1.44
MB 3.5-inch disk drive
•
Operating system support: MS Windows
®
•
CE marked and UL listed
•
Two-button mouse or equivalent
•
Monitor: A minimum of 17-inch diagonal monitor
•
Keyboard: a 104-key keyboard or equivalent