Xyp instrument problems, Power on indicator light is not on – Luminex 100 IS User Manual, Version 2.1 User Manual
Page 156
Luminex 100 IS User Manual Version 2.1
x
MAP Technology
9 - 10
PN 89-00002-00-061 Rev. A
•
Complete the following analyzer functions from the Maintenance
tab: Drain, Backflush, and Alcohol Flush.
•
Empty the waste container. Prime the system and check if sheath
fluid is dripping into the waste container. Contact Luminex
Customer Support with this information.
•
If using the filter bottom plates, adjust the sampling probe height
up to prevent the probe from puncturing the filter.
Progress Bar Indicates No
Events Being Collected
During Calibration or
Verification
•
Ensure that the Luminex 100 analyzer and XYP instrument are
connected properly.
•
Empty the waste container if it is completely filled.
•
See if the tubing from the sheath fluid and waste containers are
connected properly.
•
See if the system calibrators or controls were properly mixed
before they were dispensed into the wells.
•
See if you added the system calibrators or controls to the correct
well on the plate.
•
Check if the volume added to the well is correct.
•
Remove and sonicate or flush the XYP instrument Sample Probe
to remove possible clogs.
•
Check if the Sample Probe is properly aligned.
•
Complete the following analyzer functions from the Maintenance
tab: Drain, Backflush, and Alcohol Flush.
•
Check to see if the delta cal temperature < 3.0 degrees.
Recalibrate if greater and retry.
•
Power off the Luminex 100 analyzer, the XYP instrument, and
the PC. Then, power them back on.
•
Calibrate and verify the system.
XYP Instrument
Problems
Power On Indicator Light
Is Not On
•
Ensure that the XYP instrument is powered on.
•
Ensure that the system is plugged into an outlet with current.
•
If using a surge protector, see if the power strip is switched to the
On position.
•
Remove the cable and reconnect.
•
Check the fuses (see “Fuses” on page 8-10).