Luminex 100 analyzer, General, Optics – Luminex 100 IS User Manual, Version 2.1 User Manual

Page 43: Fluidics

background image

x

MAP Technology

The System

PN 89-00002-00-061 Rev. A

5 - 5

Luminex 100

analyzer

General

Operating temperature: 15°C to 30°C (59°F to 86°F)

Storage temperature: 0°C to 50°C (32°F to 122°F)

Humidity: 20% to 80%, noncondensing

Altitude: Operation up to 2400 m (7874 ft.) above mean sea level

Physical dimensions: 43 cm (17 inches) W x 50.5 cm (20 inches)
D x 24.5 cm (9.5 inches) H

Weight: maximum of 25 kg (60 lbs.)

UL installation category: UL Installation Category II, as defined
in Annex J of UL 61010A-1

Pollution degree: UL Pollution Degree 2, as defined in Section
3.7.3.2 of UL 61010A-1

Shipping and storage: The allowable shipping and storage
temperature and humidity ranges are 0°C to +50°C and 20-80%
noncondensing, respectively

Input voltage range: 100 - 240 Vac, 1.5 Amp, 47-63 Hz or 100 -
120 Vac, 1.4 Amp, 47-63 Hz. (refer to the serial number label on
the rear panel of the instrument)

AC inlet fuse: 3 Amp, 250 Vac, fast acting or 2 Amp, 250 Vac,
fast acting (refer to the fuse label on the rear panel of the
instrument)

Optics

Reporter laser: 532 nm, nominal output 10-16.5 mW, maximum
500 mW, frequency-doubled diode; mode of operation,
continuous wave (CW)

Classification laser: 635 nm, 8 mW, maximum output 25
mW, diode; mode of operation, continuous wave (CW)

Reporter detector: Photomultiplier tube, detection bandwidth of
565 - 585 nm

Classification detector: Avalanche photo diodes with
temperature compensation

Doublet discrimination detector: Avalanche photo diodes with
temperature compensation

Fluidics

Sheath flow rate: 90 µL ±5 µL/seconds

Cuvette: 200 micron square flow channel

Sample injection rate: 1 µL/second

Sample uptake volume: 20 - 200 µL