beautypg.com

Basic concepts – Luminex 100 IS User Manual, Version 2.1 User Manual

Page 23

background image

PN 89-00002-00-061 Rev. A

3 - 1

Basic Concepts

Background

Information

xMAP technology is a versatile system that measures soluble
analytes. The 100 IS performs simultaneous, discrete measurements
of multiple microsphere-based reactions from a single specimen
aliquot.

For more conceptual information, refer to Practical Flow

Cytometry, 3rd edition, by Howard M. Shapiro, M.D. (New York:
Wiley-Liss Inc., 1995).

Fluidics

There are two fluidic paths in the 100 analyzer. The first path
involves a syringe-driven mechanism that controls the sample
uptake. This mechanism permits small sample uptake volumes from
small reaction volumes. The syringe-driven system transports a
specified volume of sample from a sample container to the cuvette.
The sample is injected into the cuvette at a steady rate for analysis.
Following analysis, the sample path is automatically purged with
sheath buffer by the second fluidics path. This process effectively
removes residual sample within the tubing, valves, and probe.
Approximately 160 µL of sheath fluid is dispelled into each well
following sample acquisition. The second fluidics path is driven
under positive air pressure and supplies sheath fluid to the cuvette
and sample path.

Excitation

The excitation system in the Luminex 100 analyzer involves two
solid-state lasers. A reporter laser excites fluorescent molecules
bound to biological reactants at the microsphere surface, and a
classification laser excites fluorochromes embedded in the
microsphere. The lasers illuminate the xMAP microspheres as they
flow single file through the cuvette. RP1 refers to the excitation
wavelength. CL1 and CL2 refer to the dyes embedded in the
microsphere. DD refers to the channel that discriminates against
doublets based on size.

3