beautypg.com

HP 8360 User Manual

Page 68

background image

Access User Flatness Correction Menu

9.

Press

POWER

Select

Menu.

Select Delete Menu Delete All. This step insures that the

flatness array is empty.

11. Press

Leave the delete menu and return to the previous

soft key menu.

12. Select Copy List This step copies the frequency list into the

correction table in sequential order.

13. Select Freq Follow. This sets the synthesizer to CW frequency

mode to facilitate taking correction information. As you scroll
through the correction cells, the synthesizer produces the
corresponding CW frequency at 0

14. Select Enter Corr . This allows correction value entry.

15. Press

This step enables user flatness correction.

16. For 5

set the appropriate power sensor cal factor on the

power meter.

17. Use the synthesizer rotary knob to adjust for a measurement of

0.00

on the power meter. Notice that a correction value is

entered at 5

18. Use the up arrow key to increment to the next correction cell.

19. For 11

set the appropriate power sensor cal factor on the

power meter.

20. Use the synthesizer rotary knob to adjust for a measurement of

0.00

on the power meter.

21. Repeat this sequence of steps until all the frequency points have

a correction value entered.

Activate List Mode

22. Press SWEEP

Select Sweep Mode List .

23. The flatness correction array is ready to be applied to your setup.

Disconnect the power meter/sensor. The power produced at the
point where the power meter/sensor was disconnected is now
calibrated at the frequencies and power level specified above.

Getting Started Advanced