HP 8360 User Manual
Page 107
210
OUTPUT
220
ENTER
230
PRINT Id$
240
250 !
260
PRINT "Counter Used . . .
270
OUTPUT
280
ENTER
290
PRINT Id$
300
310 !
320
OUTPUT
ON"
330 !
340
350
PRINT "INPUT
360
____________
370
380 !
390
FOR Testpoint=First TO Last
400
OUTPUT
410
ENTER
420
OUTPUT
430
ENTER
440
PRINT Testpoint,Reading/lOOO
450
NEXT Testpoint
460 !
470
OUTPUT
OFF"
480
END
Program Comments. Lines 20 to 70: Declare variables and I/O paths
for instruments. I/O paths let you use a name for an instrument in
OUTPUT and ENTER
of a numeric address.
80 to 100: Assign values to the input test limits in
110 to 130: Clear the instrument HP-IB interfaces.
140 to 160: Reset each instrument to a known measurement state.
170 to 190: Print the test report title.
200 to 310: Query measurement instruments’ identifications for test
traceability.
320 to 330: Connect the source output signal to the output
terminals.
340 to 380: Print results table header.
390 to 460: This is the main measurement loop. Line 400 contains
two commands. :SOURce:VOLT sets the output level of the source.
is used to signal that the preceding command has finished
executing. To make an accurate
source output
must be allowed to settle. When the output has settled, *OPC? places
Getting Started Programming 1-77