Step Swp Pt
Trig Ext
Function Group
Menu Map
Description
Programming Codes
See Also
FREQUENCY
2
When this
is selected, the synthesizer steps to the next point
in the stepped frequency sweep when an external hardware trigger is
received. When the last frequency point is reached and continuous
sweep is selected, the next trigger causes the step sweep to return to
the start frequency. Connect the trigger signal to the TRIGGER
INPUT BNC. An asterisk next to the key label indicates that this
feature is active.
SCPI: SWEep:TRIGger:SOURce EXT
Analyzer: TS
Step Swp Menu, Sweep Mode Step
“Using Step Sweep,” in Chapter 1.
Function Group
Menu Map
Description
Programming Codes
FREQUENCY
NONE
This
activates swept frequency mode and makes the stop
frequency parameter the active function. The start/stop frequency
must be separated by at least 2 Hz in order to remain in the
frequency sweep mode. If start=stop frequency then the zero span
mode is entered.
FREQuency:STOP
FREQuency:MODE:SWEep
Analyzer: FB