l
DIAGnostics:INSTrument:PRINter:ADDRess
l
DIAGnostics:INSTrument:PRINter:ADDRess?
Sets the HP-IB address of the printer to use during some of the
calibration procedures when the synthesizer assumes HP-IB control.
*RST and power on do not effect this command. The default is 1.
The default value is set at memory initialization only.
l
DIAGnostics:IORW
Performs a write to the I/O Device number specified in the first
l
DIAGnostics:IORW?
Reads from the specified I/O device number and returns the response
data.
l
DIAGnostics:OUTPut:FAULt?
Returns a string of 16, and that are equivalent to the fault
display.
Bit 0 = PEAK
Bit 1 TRACK
Bit 2 = RAMP
Bit 3 = SPAN
Bit 4 =
Bit 5 = ADC
Bit 6 = EEROM
Bit 7 PWRON
Bit 8 = CALCO
Bit 9 PLLZERO
Bit 10 PLLWAIT
Bit 11 = FNXFER
Bit 12 = CAL YO
Bit 13 CAL MAN
Bit 14 = TMR CNFLCT
Bit 15 SEARCH
l
DIAGnostics:RESult?
Returns the following information:
where,
is one of the following:
0
Diagnosis successful.
1
Cannot diagnose; full
must be executed first.
2
No failures found-all selftests passed.
- 1
Cannot diagnose; diagnosis routine failed to isolate failure.
(software fault)
(-999 if parameter is not used, as in
Operating and Programming Reference
HP 8360