Programming Codes
CORRection:FLATness
The portion of the above command contained in can be
entered from one to 801 times. This command creates the
frequency-correction pair array similar to the front panel array.
The correction entered is at the associated frequency and
frequencies in between are determined by linear interpolation.
CORRection:FLATness?
This command queries the flatness array created with CORR:FLAT.
CORRection:ARRay[i]
The portion of the above command contained in must
be entered 1601 times. This array must contain 1601 evenly
spaced correction values. This command creates the
correction set that has no equivalent front panel entry. If this
command is used to enter flatness correction information the
CORRection:SOURce command (described below) will be set to
array. There is an array for the foreground state (i=O) and for the
background state
If is not specified, the default is the
foreground state
CORRection:ARRay[i]?
This command queries the entire
correction set.
CORRection:SOURce[i]
When the above command is set to flatness CORR:SOUR FLAT, the
array chosen is the frequency-correction pair array. When the
command is set to array CORR:SOUR ARR, the array chosen is the
1601 point correction set.
CORRection:SOURce[i]?
Queries the source of correction.
CORRection[:STATe]
Sets the switch on the user flatness correction feature. This is the
same as pressing
on the front panel.
CORRection:STATe?
Queries the condition of the internal switch.
CORRection:FLATness:POINts?
The above command returns information on how many
frequency-correction pairs were entered using the CORR : FLAT
command.
Analyzer:
NONE
Operating and Programming Reference
HP 8360
User’s Handbook