Creating a User Flatness Array Automatically, Example 1

In this example, a flatness array containing correction frequencies

from 4 to 10

at 1

intervals is created. An HP

power

meter controlled by the synthesizer through the interface bus is used
to enter the correction data into the flatness array.

For this example, refer to menu map 5, POWER.

1. The equipment setup shown in Figure 1-18 assumes that if the

setup has an external leveling configuration, the steps necessary
to correctly level have been followed. If you have questions about
external leveling refer to earlier paragraphs titled, “Externally
Leveling the Synthesizer.”

Setup Power Meter

2. Zero and calibrate the power meter/sensor.

3. Enter the appropriate power sensor calibration factors into the

power meter.

4. Enable the power meter/sensor cal factor array. For operating

information on the HP

power refer to its operating and

service manual.

5. Connect the power sensor to the point where corrected power is

desired.

HP
POYER NE

I

,

PORT

SENSOR

-

-

-

-

-

-

Figure 1-16. Creating a User Flatness Array Automatically

TER

1-34 Getting Started Advanced