beautypg.com

Hp-ib check, example program 1, Hp-ib, Example – HP 8360 User Manual

Page 118

background image

Table 1-4. Sample Synthesizer Commands (continued)

HP-IB

Command

Parameters

setting

coupled

output level

Parameter Type

Allowed Values

extended numeric

to [DB] or

Boolean

extended numeric

specified power range

RF on/off

Boolean

type of sweep

discrete

sweep time

extended numeric

to 133

or

auto sweep

Boolean

time switch

:LLIMit

fastest sweep
time

extended numeric

suffix] or

Example

This first program is to verify that the HP-IB connections and

1

interface are functional. Connect a controller to the synthesizer
via an HP-IB cable. Clear and reset the controller and type in the
following program:

10

20

30
40
50
60
70
80
90

ABORT 7
LOCAL Source
CLEAR Source
REMOTE Source

PRINT "The source should now be in REMOTE."
PRINT "Verify that the 'REMOTE' LED is
END

Run the program and verify that the REMOTE LED is lit on the
synthesizer. If it is not, verify that the synthesizer address is set to

19 and that the interface cable is properly connected.

If the controller display indicates an error message, it is possible that
the program was entered in incorrectly. If the controller accepts the
REMOTE statement but the synthesizer REMOTE LED does not
turn on, perform the operational checks as outlined in the respective

Operating and Service Manuals to find the defective device.

1-88 Getting Started Programming