Leveling point extdet, Leveling point intrnl, Leveling point – HP 8360 User Manual
Page 240
Leveling
Leveling Point
Function Group
Menu Map 1
Description
Programming Codes
See Also
ALC
This
lets you set the synthesizer to accept an external
feedback connection from a negative-output diode detector to level
power. The EXT ALC BNC is the input connection for the required
signal. An asterisk next to the key label indicates that this feature is
active.
POWer:ALC[:SOURCce]
POWer:ATTenuation:AUTO
Analyzer:
“Externally Leveling the Synthesizer,” in Chapter 1.
Leveling Point
Function Group
Menu Map
Description
Programming Codes
See Also
HP 8360
User’s Handbook
ALC
1
This
lets you set the synthesizer to level at the output of the
directional coupler located inside the synthesizer. An asterisk next to
the key label indicates that this feature is active.
SCPI:
POWer:ALC[:SOURce]
Analyzer:
Al
Operating and Programming Reference