background image

Start Sweep

Trigger Ext

Function Group

Menu Map

Description

Programming Codes

See Also

SWEEP

7

When this

is selected, the synthesizer waits for an external

hardware trigger to trigger a sweep. Connect the trigger pulse to
TRIGGER INPUT. It is activated on a TTL rising edge. An asterisk
next to the key label indicates that this feature is active.

SCPI: TRIGger:SOURce EXT
Analyzer:

Sweep Menu

Step Control

Master

Function Group

FREQUENCY

Menu Map

Description

This

lets you designate the synthesizer as the master control

in a dual synthesizer measurement system. A dual synthesizer
system (two-tone measurement system) facilitates accurate device
characterizations by providing one

reference for both

sources. This technique reduces instabilities from temperature or line

voltage fluctuations, or drift.

The synthesizers can be operated in either ramp sweep or step
sweep modes for both fixed offset and swept offset measurements.
Figure S-l shows the connections required for a two-tone system.

Operating and Programming Reference HP 8360