Function Group

Menu Map

Description

Programming Codes

See Also

INSTRUMENT STATE

NONE

This

(green) causes the synthesizer to perform a short

version of self-test, and initializes the synthesizer to a standard
starting configuration. Two states can be defined for the standard
configuration: Factory or User.

Press

at any time to test the synthesizer and restore to a

standard configuration. If the red LED adjacent to THE [PRESET)
KEY (labeled INSTR CHECK) t y

s a on after preset, the synthesizer

failed self-test; refer to “Troubleshooting Manual”.

Cycling power with the POWER switch does not have the same

effect as presetting the synthesizer. Cycling power causes the
synthesizer to display the programming language, the HP-IB address,

and the firmware revision date. After the synthesizer displays this
data, it restores its configuration to the state before power was
turned off.

SCPI:

SYSTem:PRESet[:EXECute]

Analyzer:

IP

Preset

Mode

Factory,

Mode

User

“Changing the Preset Parameters,” in Chapter 1.
“Programming Typical Measurements,” in Chapter 1.

Operating and Programming Reference

HP 8360

User’s Handbook