Preset – HP 8360 User Manual
Page 282
Function Group
Menu Map
Description
Programming Codes
See Also
INSTRUMENT STATE
NONE
This
(green) causes the synthesizer to perform a short
version of self-test, and initializes the synthesizer to a standard
starting configuration. Two states can be defined for the standard
configuration: Factory or User.
Press
at any time to test the synthesizer and restore to a
standard configuration. If the red LED adjacent to THE [PRESET)
KEY (labeled INSTR CHECK) t y
s a on after preset, the synthesizer
failed self-test; refer to “Troubleshooting Manual”.
Cycling power with the POWER switch does not have the same
effect as presetting the synthesizer. Cycling power causes the
synthesizer to display the programming language, the HP-IB address,
and the firmware revision date. After the synthesizer displays this
data, it restores its configuration to the state before power was
turned off.
SCPI:
SYSTem:PRESet[:EXECute]
Analyzer:
IP
Preset
Mode
Factory,
Mode
User
“Changing the Preset Parameters,” in Chapter 1.
“Programming Typical Measurements,” in Chapter 1.
Operating and Programming Reference
HP 8360
User’s Handbook