beautypg.com

Dblr amp menu, Dblr menu – HP 8360 User Manual

Page 188

background image

Dblr

Menu

Function Group

POWER

Menu Map

Description

This

accesses the doubler amp mode softkeys. These

are applicable to instrument models with a doubler installed. The
doubler has an integral amplifier whose operation is controlled by the
instrument firmware. Its use depends on the frequency of operation
and on the calibration constants set at the factory. The instrument
defaults after preset to this automatic mode of operation which is
the specified operation.

in this menu will allow you to turn

the doubler amplifier always on or always off. These two modes are

unspecified operation for instruments with a doubler installed. These

have no effect on instruments without a doubler.

Doubler Amp Mode AUTO

Sets the doubler amp mode to
AUTO. This is the default after
preset and must be used for specified
performance.

Doubler Amp Mode On

Turns the doubler amplifier on
regardless of the frequency of
operation. Using this mode results in

unspecified performance.

Doubler

Mode Off

Turns the doubler amplifier off
regardless of the frequency of
operation. Using this mode results in

unspecified performance.

Programming Codes

SCPI:

NONE

Analyzer:

NONE

See Also

listed above.

HP 8360

User’s Handbook

Operating and Programming Reference D-l