Programming codes – HP 8360 User Manual
Page 244
List Menu
To remove a frequency point and its associated offset value and dwell
time, use the delete menu (Delete
To remove an
entire frequency list, use the delete me
(Delete All
Editing ALC Offset and Dwell Time
Once a frequency point has been entered, You can assign an ALC
offset and a dwell time value. Use either the Enter List Power
or Global Offset
to enter offset values. Use either the
Enter List Dwell or Global Dwell.
to enter dwell time
values.
The editing
of this menu are not accessible over HP-IB.
Frequency lists to be loaded over HP-IB must first be created in the
controlling program and then downloaded in their entirety to the
synthesizer.
Programming Codes
{
LIST:[POWer]:CORRection
{
In the above three commands, the entries contained in can be
repeated between 1 to 801 times.
LIST:FREQuency:POINts?
LIST[:POWer]:CORRection:POINts?
In the above three commands, the synthesizer responds with the
number of points for the named parameter that are in the list array.
If a particular list is shorter than another, an error is generated upon
execution. An exception is made for the case where the shorter list
is of length 1. In this case, the list of length 1 is treated as though
it were a list of equal length, with all values the same. At
all
lists for the current state are cleared and reset to a single value.
Analyzer:
NONE
(SAVE), Sweep Mode List
“Creating and Using a Frequency List,” in Chapter 1.
HP 8360
User’s Handbook
Operating and Programming Reference