Hint – HP 8360 User Manual
Page 56
External Leveling Used With the Optional Step Attenuator
Some external leveling applications require low output power
from the synthesizer. The synthesizer automatically uncouples the
attenuator from the ALC system for
external leveling points.
Press
(
P O W E R L E V E L
).
Note the display. It shows:
0
POWER LEVEL: 0.00
For example, leveling the output of a 30
gain amplifier to a level
of -10
requires the output of the synthesizer to be around
-40
when leveled. At some frequencies this level is beyond the
range of the ALC modulator alone. If so, the LOW UNLVLED warning
message is displayed. Inserting 40
of attenuation results in an
ALC level of 0
which is well within the range of the ALC. At
20
attenuation is a better choice as it results in an ALC
level of -10
This gives a margin for AM or other functions
that vary the power level.
For optimum display accuracy and minimum noise, the ALC
level should be greater than -10
This is achieved by using
attenuation equal to the tens digit of output power. Example:
desired output power = -43
use:
40
, ALC -3
1. Press POWER
Select Set
Hint
To obtain flatness corrected power refer to “Optimizing Synthesizer
Performance, Creating and Applying the User Flatness Correction
Array,” later in this section.
Getting Started Advanced